User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: EICO 1171:Resistance Deacade Box DATE: 09-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 55 CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P U F W 1.003 HEAD {* STORE ZERO OFFSET *} 1.004 DISP Set all decades to zero (to measure the zero offset). 1.005 DISP [32] Connect the 1281 and the UUT as follows: 1.005 DISP [32] 1281 TO EICO 1171 1.005 DISP [32] INPUT HI ......... DECADE HI 1.005 DISP [32] INPUT LO ......... DECADE LO 1.005 DISP [32] INPUT I+ ......... DECADE HI 1.005 DISP [32] INPUT I- ......... DECADE LO 1.006 IEEE [@1281]OHMS AUTO, FWR 1.007 IEEE [@1281][D2000]RDG?[I] 1.008 RSLT =Zero Offset =[mem] ohms 1.009 MATH M[3] = MEM 1.010 HEAD {* DECADE 1 Ohm/STEP *} 1.011 MATH M[1] = 1.0000 1.012 TOL .5% .1000U 1.013 CALL Sub Decade 9 Steps CAL VER 1.014 JMP 2.001 1.015 EVAL dummy line 2.001 HEAD {* DECADE 10 Ohm/STEP *} 2.002 MATH M[1] = 10.000 2.003 TOL .5% .1000U 2.004 CALL Sub Decade 9 Steps CAL VER 2.005 JMP 3.001 2.006 EVAL dummy line 3.001 HEAD {* DECADE 100 Ohm/STEP *} 3.002 MATH M[1] = 100.000 3.003 TOL .5% .1000U 3.004 CALL Sub Decade 9 Steps CAL VER 3.005 JMP 4.001 3.006 EVAL dummy line 4.001 HEAD {* DECADE 1k Ohm/STEP *} 4.002 MATH M[1] = 1000.00 4.003 TOL .5% .1000U 4.004 CALL Sub Decade 9 Steps CAL VER 4.005 JMP 5.001 4.006 EVAL dummy line 5.001 HEAD {* DECADE 10k Ohm/STEP *} 5.002 MATH M[1] = 10000.00 5.003 TOL .5% .1000U 5.004 CALL Sub Decade 9 Steps CAL VER 5.005 END